Basic function principles
![]() | RMS value (rms) specifications All AC value specifications in this documentation such as RMS specifications (rms) refer to a 50/60 Hz 3-phase mains network with a sinusoidal waveform. |
The EL3773 has 6 analog input channels which are measured simultaneously in their respective measuring ranges. The properties are discussed below in the order of the data processing.
Input filter
Each channel has a 5th order low-pass filter with Bessel characteristic which can be parameterized via the CoE setting 0x8000:15. The corresponding anti-aliasing filters are placed up or downstream and are adapted automatically.
The filter setting of channel 1 also applies to channels 2 to 6; the corresponding CoE object cannot be written in these channels. All 6 input channels are thus subject to the same filter characteristic.
The input filter cannot be deactivated.
Fig.95: Data flow diagram input filterThe following steps can be selected:
- 200 Hz
- 500 Hz
- 1 kHz
- 1.5 kHz
- 2.5 kHz
- 5 kHz
- 15 kHz
Fig.96: Filter setting via the CoE object "Filter settings" (Index 0x8000:15)The amplitude curve can be represented accordingly:
Fig.97: Amplitude curve![]() | Changes in the CoE directory In case of changes to the CoE default parameters, it is essential that corresponding values are entered in the start-up list, so that in the case of exchange the EL3773 operates again as foreseen in the application. |