Measured value filter

Each measuring axis has an integrated digital filter. You can enable and configure the filter individually for each measuring axis.

Measured value filter 1:

Recommendation: use the "Filter Designer"

Use the TE1310 | "TwinCAT 3 Filter Designer" for a simplified configuration of the filters, see https://www.beckhoff.com/te1310.

Manual configuration

If you do not use the "TwinCAT 3 Filter Designer", you must configure the filters manually using CoE parameters. This is particularly recommended for simple predefined filters.

Set the type of filter in the CoE parameter "Filter Settings" of the respective measuring axis:

Measuring axis

Parameter "Filter Settings"

Acc X

8000:15

Acc Y

8010:15

Acc Z

8020:15

Gyro X

8030:15

Gyro Y

8040:15

Gyro Z

8050:15

The following filter types are available there:

Value

Name

Comment

0

None

Filter disabled.

1

FIR Notch 50 Hz

Notch filter as 50 Hz hum filter.

2

FIR Notch 60 Hz

Notch filter as 60 Hz hum filter.

19dec

IIR Butterw. LP 5th Ord. 25Hz

5th order Butterworth low-pass filter with a cut-off frequency of 25 Hz.

32dec

User Defined FIR Filter

Recommendation: only use this filter type with the "TwinCAT Filter Designer", see above.

33dec

User Defined IIR Filter

Recommendation: only use this filter type with the "TwinCAT Filter Designer", see above.

34dec

User Defined Average Filter

Moving average filter. Set the number of averaged measured values (samples) in parameter 80n1:01 "Filter Coefficient 1".

Notch filter curve, example for “FIR Notch 50 Hz”

Measured value filter 2: