Measuring ranges

You can set the measuring ranges for acceleration and angular velocity in parameter F800:11 "Range". Factory setting: "+/-2000dps, +/-16g".
This parameter applies to all six measuring axes. An individual setting for individual axes is not possible.

Measuring ranges 1:

The setting of the measuring ranges also influences the following factors:

Setting via numerical values

You will need the following information if you do not set the measuring ranges in TwinCAT via the dialog box.

The angular velocity measuring range and the acceleration measuring range are each coded in a hexadecimal digit ("Nibble") in parameter F800:11 "Range":

Measuring ranges 2:

Possible values for the angular velocity measuring range:

Value

Angular velocity measuring range

0

+/- 2000dps

1

+/- 1000dps

2

+/- 500dps

3

+/- 250dps

4

+/- 125dps

5

+/- 62.5dps

6

+/- 31.25dps

7

+/- 15.625dps

Possible values for the acceleration measuring range:

Value

Acceleration measuring range

0

+/- 16g

1

+/- 8g

2

+/- 4g

3

+/- 2g