Restoring the delivery state - Implementation in ELM3xxx

Notice

Function availability

This function is available in
- ELM3002 from FW08
- ELM3004 from FW10
- ELM3102-00x0 from FW09
- ELM3102-01x0 from FW04
- ELM3104 from FW09
- ELM324x from FW02

According to the documentation chapter Restoring the delivery state, two reset mechanisms "load" and "wipe" are available, which are implemented in ELM3xxx as follows:

Data

PW protection

load

wipe

disabled

enabled

disabled

enabled

User parameters

0x80n0: PAI Settings Ch.n

reset

reset

reset

reset

0x80n1: PAI Filter 1 Settings Ch.n

reset

reset

reset

reset

0x80n3: PAI Filter 2 Settings Ch.n

reset

reset

reset

reset

0x80n6: PAI Scaler Settings Ch.n

reset

reset

reset

reset

0x80n7: PAI RTD Settings Ch.n

reset

reset

reset

reset

0x80nE: PAI User Calibration Data Ch.n

reset

keep

reset

reset

0xF800: PAI Settings Device

reset

reset

reset

reset

0xF009: Password protection

keep

keep

reset

reset

SM Synchron (default) / FreeRun

0x1C32:01 SM output parameter, Sync mode

keep

keep

reset

reset

0x1C33:01 SM input parameter, Sync mode

keep

keep

reset

reset