Measurement error/measurement deviation/measurement uncertainty, output uncertainty
![]() | Analog output The following information also applies analogously to the output end value of analog output devices. |
The relative measuring error as a specification value of a Beckhoff analog device is specified in % of the nominal FSV (output end value) and calculated as the quotient of the numerically largest probable deviation from the true measured value (output value) with respect to the FSV (output end value):
It should be noted here that the "true measured value" cannot be determined with infinite accuracy either, but can only be determined via reference devices with a higher expenditure of technology and measuring time and thus a significantly lower measurement uncertainty.
The value therefore describes the result window in which the measured value determined by the device under consideration (Beckhoff analog device) lies with a very high probability in relation to the "true value". Thus, colloquially, this is a "typical" value (typ.); this expresses that the vast statistical majority of values will be within the specification window, but in rare cases there may/will be deviations outside the window.
For this reason, the term "measurement uncertainty" has become established for this window, since "error" is now used to refer to known disturbance effects that can generally be systematically eliminated.
The uncertainty of measurement must always be considered in relation to potential environmental influences:
- invariable electrical channel properties such as temperature sensitivity,
- variable settings of the channel (noise via filters, sampling rate, ...).
Measurement uncertainty specifications without further operational limitation (also called "service error limit") can be assumed as a value "over everything": entire permissible operating temperature range, default setting, etc.
The window is always to be understood as a positive/negative span with "±", even if occasionally indicated as a "half" window without "±".
The maximum deviation can also be specified directly.
Example: measuring range 0...10 V (FSV = 10 V) and measurement uncertainty < ± 0.3%FSV → the expected maximum usual deviation is ± 30 mV in the permissible operating temperature range.
![]() | Lower measurement uncertainty possible If this specification also includes the temperature drift, a significantly lower measuring error can usually be assumed in case of a constant ambient temperature of the device and thermal stabilization after a user calibration. |