Installation instructions for cascaded connection
Observe the following instructions for using cascaded analog inputs!
- 1. All channels of a terminal must be used consecutively; gaps in usage, that is, unused terminal connection points (points 2, 3, 4, 5, 6, 7, and 8 in the connection diagram), are not recommended, as they can lead to significant measuring errors
- 2. The terminal is designed for the measurement of a high number of channels. The wiring between terminals must be carried out as shown in the following figure.
- 3. If the number of cells to be measured is less than the number of available channels (n terminals × 8), unused channels will result at the first/last terminal. Unused channels at the top/bottom end of the terminal display random values. It is therefore recommended to connect the unused channels electrically to the last used pole using wire jumpers!
- 4. The cascaded layout of the channels results in a channel-by-channel differential measurement; see detailed diagram: “Cell 3” has a connection to terminal connection point 3 via wire 3b and wire 3, and simultaneously has a connection to terminal connection point 4 via wire 4a and wire 4.
This means that each cell – and thus each measuring channel from the user’s perspective – is connected to 2 terminal connection points!
Therefore, when it comes to wire breaks in the system, the following distinctions must be made
- Wire break at a terminal connection point, e.g., 3, 4, 5…:
An open terminal or a wire break in the supply line (3, 4…) results in an incorrect measured value or even an out-of-range reading on both affected measuring channels, thereby affecting 2 measured values. The terminal detects at least one “Overrange” event, and the channel enters the “Error” state (note the filter times; with the factory setting (IIR8 filter), this detection takes approximately 3 seconds). - Wire break in a cell stack connection, e.g., 3b or 4a:
The entire stack is interrupted; the affected cell can no longer be measured; the affected channel/measured value will display a value in the range of approximately 2 times the cell voltage due to the open circuit, but will not show an “Error” on a per-channel basis. This error condition can only be detected at the stack level by the measurement of the stack voltage.